Closed ESR System


ESR fast detector is made up of plastic frame with a graduation scale indicating ESR Values and a stainless steel motherboard. It is suitable for the ESR tests for subjects of different gender and ages, wtih higher related coefficient comparing wtih the Westergren method.


Intended Use

• ESR fastdeteclor Is designed to match IMPROVACUTER·
•ESR rubes(specIftcatIon: 1.6mV 9x120mm) for testing the (ESR). which Is a non-specltic assay Item.
• It Is safe, reliable and easy to read with clear scales.


Installation

• ESR fast detector should be placed stably without wobble and draught.
• It should be kept erect with 8 slope angle less than 2°


Specimen Collection and Processing


1. Make use of ESR tube and needles to obtain the blood specimen by venipuncture.
2. Gently Invert ESR tube at least 8-10 times. Avoid of fierce shake to destroy the Erythrocytes.
3. Repeat before Inserting Tube into Stand.
4. Insert ESR tube into Stand.
5. Align zero level of scaie to bottom of meniscus.
6. Set timer and read erythrocyte level after 30 min.
7. Discard ESR tubes without opening.